Radiation Dose Estimates for Kr-81m Injections
| |
Estimated
Radiation Dose (mGy/MBq) |
| ORGAN |
mGy/MBq |
rad/mCi |
| |
|
|
| Adrenals |
4.5E-06 |
1.7E-05 |
| Brain |
9.7E-08 |
3.6E-07 |
| Breasts |
4.5E-06 |
1.7E-05 |
| Gallbladder Wall |
1.7E-06 |
6.4E-06 |
| LLI Wall |
8.9E-08 |
3.3E-07 |
| Small Intestine |
3.0E-07 |
1.1E-06 |
| Stomach |
2.8E-06 |
1.0E-05 |
| ULI Wall |
4.3E-07 |
1.6E-06 |
| Heart Wall |
1.9E-04 |
7.1E-04 |
| Kidneys |
1.2E-06 |
4.4E-06 |
| Liver |
3.8E-06 |
1.4E-05 |
| Lungs |
5.1E-05 |
1.9E-04 |
| Muscle |
1.6E-06 |
6.0E-06 |
| Ovaries |
1.2E-07 |
4.4E-07 |
| Pancreas |
4.3E-06 |
1.6E-05 |
| Red Marrow |
2.1E-06 |
7.6E-06 |
| Bone Surfaces |
2.2E-06 |
8.3E-06 |
| Skin |
6.7E-07 |
2.5E-06 |
| Spleen |
2.4E-06 |
8.7E-06 |
| Testes |
1.4E-08 |
5.2E-08 |
| Thymus |
1.3E-05 |
5.0E-05 |
| Thyroid |
1.1E-06 |
4.0E-06 |
| Urinary Bladder Wall |
5.1E-08 |
1.9E-07 |
| Uterus |
1.1E-07 |
4.2E-07 |
| |
|
|
| Effective Dose Equivalent |
2.0E-05 mSv/MBq |
7.5E-05 rem/mCi |
Residence times based on a model in which activity clears
from the blood,
to the right heart, lungs, left heart, and back to blood, with
95% clearance
of blood from lungs on each pass. Half times for transport are
1 minute in
heart chambers, 5 minutes in lung, and 3 minutes in the body.
(No patient or animal data used).
Estimate calculated using phantom of Cristy & Eckerman
(Report ORNL/TM-8381/V1 & V7)